Innovative air flow system, which enables high temperature uniformity and stability
Simple microprocessor controller with a LED display
Timer or continuous operation
PID temperature controller
Visual and sound alarm in the event of failure
LED indication:
Operation of the heaters
Operation of the fan
Programme in operation
Programme completed
Programme stopped with STOP
Temperature sensor error
Display temperature resolution 0.1°C
Temperature stability ± 0.1°C
Adjustable exhaust system with a lever on the doors
Double protection system against high temperatures
when the temperature is 5°C above the set level, the operation of the heating elements is disabled
mechanical over temperature shut off device
The interiorcasing is made of stainless steel AISI 304
The exterior is made of aluminized sheet, powder coated (RAL 7035)
Two stainless steel wire shelves
Interlock with the possibility of locking (keys added)
RS – 232 communication port
Programme for the temperature probe calibration
Instructions for working with the electrical layout and the CE declaration of conformity
Type
Volume
Internal dimensions (WxHxL) mm
External dimensions (WxHxL) mm
Wattage
SP – 45
44 L
355 x 335 x 370
590 x 560 x 600
230V~50/60 Hz 1,25 kW
SP – 105
105 L
490 x 500 x 430
725 x 725 x 660
230V~50/60 Hz 1,65 kW
SP – 160
163 L
650 x 560 x 450
855 x 785 x 680
230V~50/60 Hz 2,25 kW
SP – 190
190 L
600 x 615 x 515
835 x 840 x 745
230V~50/60 Hz 2,25 kW
SP – 210
210 L
700 x 600 x 500
935 x 825 x 730
230V~50/60 Hz 2,25 kW
SP – 250
250 L
500 x 1000 x 500
735 x 1230 x 720
230V~50/60 Hz 3,3 kW
SP – 265
264 L
600 x 800 x 550
835 x 1025 x 780
230V~50/60 Hz 3,25 kW
SP – 440
440 L
1000 x 800 x 550
1235 x 1025 x 735
3x400V~50/60 Hz 3,7 kW
SP – 210
210 L
700 x 600 x 500
935 x 825 x 730
230V~50/60 Hz 2,25 kW
SP – 910
910 L
1000 x 1300 x 700
1280 x 1975 x 875
3x400V~50/60 Hz 9,5 kW
SP – 1300
1288 L
1060 x 1430 x 850
1340 x 2100 x 990
3x400V~50/60 Hz 10,5 kW
Laboratory dryers, sterilizers –'SP' series and the 'C' series controller
Standard equipment:
Temperature range Tambient + 5°C ÷ 300°C
Forced air circulation with a fan
Innovative air flow system, which enables high temperature uniformity and stability
Microprocessor controller with an LCD display – fully programmable
10 automatic programmes
Up to 10 steps in each programme
Fan speed adjustment
Start-up (at a specified time and day)
… other advanced functions
Timer or continuous operation
PID temperature controller
Visual and sound alarm in the event of failure
LED indication:
Operation of the heaters
Operation of the fan
Programme in operation
Programme completed
Programme stopped with STOP
Temperature sensor error
Display and temperature setting resolution 0.1°C
Temperature stability ±0.1°C
An adjustable exhaust system with a lever on the doors
Double protection system against high temperatures
when the temperature is 5°C above the set level, the operation of the heating elements is disabled
mechanical over temperature shut off device
The interior casing is made of stainless steel AISI 304
The exterior is made of aluminized sheet, powder coated RAL 7035
Two stainless steel wire shelves
RS – 232 communication port
Interlock with the possibility of locking (keys added)
Programme for the temperature probe calibration
Instructions for working with the electrical layout and the CE declaration of conformity
Type
Volume
Internal dimensions (WxHxL) mm
External dimensions (WxHxL) mm
Wattage
SP – 45 C
44 L
355 x 335 x 370
590 x 560 x 600
230V~50/60 Hz 1,25 kW
SP – 105 C
105 L
490 x 500 x 430
725 x 725 x 660
230V~50/60 Hz 1,65 kW
SP – 160 C
163 L
650 x 560 x 450
855 x 785 x 680
230V~50/60 Hz 2,25 kW
SP – 190 C
190 L
600 x 615 x 515
835 x 840 x 745
230V~50/60 Hz 2,25 kW
SP – 210 C
210 L
700 x 600 x 500
935 x 825 x 730
230V~50/60 Hz 2,25 kW
SP – 250 C
250 L
500 x 1000 x 500
735 x 1230 x 720
230V~50/60 Hz 3,3 kW
SP – 265 C
264 L
600 x 800 x 550
835 x 1025 x 780
230V~50/60 Hz 3,25 kW
SP – 440 C
440 L
1000 x 800 x 550
1235 x 1025 x 735
3x400V~50/60 Hz 3,7 kW
SP – 910 C
910 L
1000 x 1300 x 700
1280 x 1975 x 875
3x400V~50/60 Hz 9,5 kW
SP – 1300 C
1288 L
1060 x 1430 x 850
1340 x 2100 x 990
3x400V~50/60 Hz 10,5 kW
Laboratory dryers, sterilizers - 'S' series with natural air circulation
Standard equipment:
Temperature range Tambient + 5°C ÷ 300°C
Natural circulation without the fan
Innovative air flow system, which enables high temperature uniformity and stability
Simple microprocessor controller with a LED display
Timer or continuous operation
PID temperature controller
Visual and sound alarm in the event of failure
LED indication:
Operation of the heaters
Programme in operation
Programme completed
Programme stopped with STOP
Temperature sensor error
Display temperature resolution 0.1°C
Temperature stability ±0.1°C
An adjustable exhaust system with a lever on the doors
Double protection system against high temperatures
when the temperature is 5°C above the set level, the operation of the heating elements is disabled
mechanical over temperature shut off device
The interior casing is made of stainless steel AISI 304
The exterior is made of aluminized sheet, powder coated (RAL 7035)
Two stainless steel wire shelves
RS – 232 communication port
Interlock with the possibility of locking (keys added)
Programme for the temperature probe calibration
Instructions for working with the electrical layout and the CE declaration of conformity
Type
Volume
Internal dimensions (WxHxL) mm
External dimensions (WxHxL) mm
Wattage
S – 50
50 L
355 x 335 x 415
590 x 560 x 600
230V~50/60 Hz 1,2 kW
S – 115
116 L
490 x 500 x 475
725 x 725 x 660
230V~50/60 Hz 1,6 kW
S – 205
206 L
600 x 615 x 560
835 x 840 x 745
230V~50/60 Hz 2,2 kW
S – 285
285 L
600 x 800 x 595
835 x 1025 x 780
230V~50/60 Hz 3,25 kW
S – 475
476 L
1000 x 800 x 595
1235 x 1025 x 735
3x400V~50/60 Hz 3,7 kW
Laboratory dryers, sterilizers – 'S' series with natural air circulation and a 'C' series controller
Standard equipment:
Temperature range Tambient + 5°C ÷ 300°C
Natural air circulation without the fan
Innovative air flow system, which enables high temperature uniformity and stability
Microprocessor controller with an LCD display – fully programmable
10 automatic programmes
Up to 10 steps in each programme
Start-up (at a specified time and day)
… other advanced functions
Timer or continuous operation
PID temperature controller
Visual and sound alarm in the event of failure
LED indication:
Operation of the heaters
Programme in operation
Programme completed
Programme stopped with STOP
Temperature sensor error
Display and temperature setting resolution 0.1°C
Temperature stability ±0.1°C
An adjustable exhaust system with a lever on the doors
Double protection system against high temperatures
when the temperature is 5°C above the set level, the operation of the heating elements is disabled
mechanical over temperature shut off device
The interior casing is made of stainless steel AISI 304
The exterior is made of aluminized sheet, powder coated (RAL 7035)
Two stainless steel wire shelves
RS – 232 communication port
Interlock with the possibility of locking (keys added)
Programme for the temperature probe calibration
Instructions for working with the electrical layout and the CE declaration of conformity
Type
Volume
Internal dimensions (WxHxL) mm
External dimensions (WxHxL) mm
Wattage
S – 50 C
50 L
355 x 335 x 415
590 x 560 x 600
230V~50/60 Hz 1,2 kW
S – 115 C
116 L
490 x 500 x 475
725 x 725 x 660
230V~50/60 Hz 1,6 kW
S – 205 C
206 L
600 x 615 x 560
835 x 840 x 745
230V~50/60 Hz 2,2 kW
S – 285 C
285 L
600 x 800 x 595
835 x 1025 x 780
230V~50/60 Hz 3,25 kW
S – 475 C
476 L
1000 x 800 x 595
1235 x 1025 x 735
3x400V~50/60 Hz 3,7 kW
Simple dryers, sterilizers - 'EASY' series dry
Standard equipment:
Temperature range Tambient + 5°C ÷ 300°C
Forced air circulation with a fan
Simple digital controller with LED display
PID temperature regulator
Visual and sound alarm in the event of failure
Timer function – one stage
Display temperature resolution 0.1°C
Adjustable exhaust
The interior casing is made of stainless steel (AISI 304)
The exterior is made of aluminized sheet, powder coated (RAL 7035)
Two stainless steel wire shelves
RS – 232 communication port
Interlock with the possibility of locking (keys added)
Programme for the temperature probe calibration
Instructions for working with the electrical layout and the CE declaration of conformity
Functionality:
Sterilization with hot air
Material drying
Heating medium or industrial masses
Advantages:
Very simple to use
Perfect balance between the exterior dimensions and used volume
Inner rounded edges of the chamber enable a simple and quality cleaning
Light weight of the equipment
Easily adjustable shelf height
Type
Volume
Internal dimensions (WxHxL) mm
External dimensions (WxHxL) mm
Wattage
SP - 25 EASY
25 L
250 x 350 x 300
360 x 470 x 520
230V~50/60Hz 0,9 kW
SP - 55 EASY
55 L
400 x 400 x 350
510 x 520 x 570
230V~50/60Hz 0,9 kW
SP - 120 EASY
120 L
600 x 500 x 400
710 x 620 x 620
230V~50/60Hz 1,6 kW
SP - 240 EASY
240 L
600 x 800 x 500
710 x 920 x 720
230V~50/60Hz 1,6 kW
Additional equipment:
Calibration with a report (10 points 3 temperatures)
Calibration with a report (1 point 1 temperature)
An additional shelf for all models
Ventilation unit for additional extraction
Special extra features depending on the client's needs
Laboratory varnish dryers with special protection systems
The LSP Dryers are designed to assist the industrial processes of varnish drying using the processes of COIL COATING, HOT AIR CYCLE, etc. We recommend them to paint and varnish manufacturers for material testing and optimum setting of working parameters with larger dryers.
All dryers have a built-in programme for rapid temperature probe calibration.
Standard equipment:
Temperature range Tambient + 5°C ÷ 300°C
Forced air circulation with a fan
Forced circulation of fresh air – safety function, when the flow is suspended, the operation of heaters is shut down
Fresh air flow control – safety function, when the flow is suspended, the operation of the heaters is shut down
Flow control of inner air
Microprocessor controller with an LCD display – fully programmable
10 automatic programmes
Up to 10 steps in each programme
Start-up (at a specified time and day)
… other advanced functions
Timer or continuous operation
PID temperature controller
Visual and sound alarm in the event of failure
LED indication:
Operation of the heaters
Operation of the fan
Programme in operation
Programme completed
Programme stopped by STOP
Error
Display and temperature setting resolution 0.1°C
Temperature stability ±0.1°C
Double protection system against the high temperatures
when the temperature is 5°C above the regulated level, the operation of the heaters is shut down,
mechanical safety thermostat
The interior casing is made of stainless steel AISI 304
The exterior is made of aluminized sheet, powder coated (RAL 7035)
Two stainless steel wire shelves
RS – 232 communication port
Interlock with the possibility of locking (keys added)
Programme for the temperature probe calibration
Instructions for working with the electrical layout and the CE declaration of conformity
Type
Internal dimensions (WxHxL) mm
External dimensions (WxHxL) mm
Wattage
LSP-45 C
350 x 350 x 350
940 x 845 x 960
230V~50/60Hz 1,6 kW
LSP-105 C
490 x 500 x 430
870 x 785 x 615
230V~50/60Hz 3,1 kW
LSP-190 C
600x615x515
980x900x700
3x400V~50/60Hz 4,6 kW
LSP-210 C
580 x 600 x 600
1000 x 1550 x 1100
3x400V~50/60Hz 12,0 kW
LSP-215 C
LSP-252 C
High Temperature Chambers - Dryers
Standard equipment:
Temperature range Tambient + 5°C - 500°C (600°C)
Forced air circulation with a fan
Innovative air flow system, which enables high temperature uniformity and stability
Microprocessor controller with an LCD display – fully programmable
10 automatic programmes
Up to 10 steps in each programme
Fan speed adjustment
Start-up (at a specified time and day)
Other advanced functions
Timer or continuous operation
PID temperature controller
Visual and sound alarm in the event of failure
Display and temperature setting resolution: 1°C
Temperature stability ± 1°C
Adjustable exhaust system
Double protection system against the elevated temperature
when the temperature is 5°C above the regulated, the operation of the heaters elements is shut down,
mechanical safety thermostat
The exterior is made of aluminized sheet, powder coated (RAL 7035)
Two stainless steel wire shelves
RS – 232 communication port
Programme for the temperature probe calibration
Strong construction
Ceramic insulation with an air pocket to prevent the heating of the exterior casing
The instructions for working with the electrical layout and the CE declaration of conformity
Type
Internal dimensions (WxHxL) mm
External dimensions (WxHxL) mm
Wattage
SP – 60 C FIRE
400 x 400 x 390
950 x 670 x 780
3x400V~50/60Hz 3,2 kW
SP – 60 C FIRE (230V 1N)
400 x 400 x 390
950 x 670 x 780
230V~50/60Hz 2,8 kW
SP – 190 C FIRE
600 x 600 x 530
950 x 845 x 960
3x400V~50/60Hz 9,2 kW
SP – 320 C FIRE
825 x 525 x 820
1540 x 1620 x 1340
3x400V~50/60Hz 20 kW
Optional additional equipment
Calibration with a report (10 points, 3 temperatures)
Additional wire shelf
Additional shelf made of perforated sheet
Access port with plugs (S 40mm , S 50mm , S 90mm)
The interior is constructed with the AISI 316 material
Observation glass on the doors and a light in the chamber
Printer for recording the procedure
Software equipment for monitoring and recording the measured values on a PC